Conformity Assessment Routes Under MDR 2017/745

Sep 22, 2021 | Acorn Regulatory News, Medical Devices

As part of our ongoing review at the changes introduced with MDR, in this article, we focus on Conformity Assessment routes. We provide a revised examination of the changes under MDR and the assessment routes for different device classifications.

Increase in Requirements

Under MDR 2017/745 , Conformity Assessment routes have not changed significantly compared to MDD 93/42/EEC.  However, the timeline to CE marking might now be longer, particularly if there is involvement of expert bodies e.g. the MDCG (Medical device co-ordination group).

Notified body involvement is required for all Class IIa, IIb and III devices and may be required for some specific Class I devices. MDD 93/42/EEC referenced 13 Essential Requirements and AIMDD 90/385/EEC referenced 16 Essential Requirements. The MDR references 23 Safety & Performance Requirements.

Medical Devices are required to comply with the relevant GSPRS (Annex I). Compliance with EN harmonised standards and Common specifications (CS) presumes compliance with the GSPR’s.

Summary of Conformity Assessment Routes under MDR 2017/745

Class I

1. Device Classification:

Class I (with the exception of Class Im (measuring), Class Is (sterile) and Class Ir (reusable)

Conformity Assessment Route:

Self-certification in addition to Maintenance of Technical Documentation according to Annex II & III

2. Device Classification:

Class Im (measuring), Class Is (sterile), Class Ir (reusable)

Conformity Assessment Route:

  • Quality Management System required to control production (Annex XI Part A) or to control special characteristics (Annex IX Chapters I & III)
  • Notified Body involvement required to assess QMS

Class IIa

1. Device Classification:

Class IIa

Conformity Assessment Route:

  • Annex IX (Full QMS) or
  • Annex XI
  • (Part A – Production Quality Assurance & Part B – Product Verification) in addition to Technical Documentation as per Annex II & III

2. Device Classification:

Class Im (measuring), Class Is (sterile), Class Ir (reusable)

Conformity Assessment Route:

  • Quality Management System required to control production (Annex XI Part A) or to control special characteristics (Annex IX Chapters I & III)
  • Notified Body involvement required to assess QMS

*Note – Conformity assessment under Annex IX (Full QMS) will involve:

  • NB audit of QMS.
  • NB assessment of the Technical Documentation/Design Dossier.
  • NB audit of Production Quality Assurance for the ability to produce and test the device.

Class IIb

Class IIb devices categories are:

  1. Class IIb active device intended to administer and or remove medicinal products, body liquids or other substances to or from the body.
  2. Class IIb implantable (Well Established Technologies (WET))
  3. Class IIb non-implantable

1. Device Classification:

Class IIb active device

Conformity Assessment Route:

  • Annex IX, QMS – Chapters I, III
  • Technical documentation
  • Annex X – Type examination
  • Annex XI
  • Part A – Product Quality Assurance
  • Part B – Product Verification
  • Clinical Evaluation
  • Annex IX Section 5/Annex X Section 6
  • Declaration of Conformity

2. Device Classification:

Class IIb Implantable (WET))and non-implantable (non WET)

Conformity Assessment Route:

  • Annex IX, QMS – Chapters I,II & III
  • Technical documentation
  • Annex X – Type examination
  • Annex XI
  • Part A – Product Quality Assurance
  • Part B – Product Verification
  • Declaration of Conformity

3. Device Classification:

Class IIb

Conformity Assessment Route:

  • Annex IX, QMS – Chapters I & III
  • Technical documentation
  • Annex X – Type examination
  • Annex XI
  • Part A – Product Quality Assurance
  • Part B – Product Verification
  • Declaration of Conformity

Class III

1. Device Classification:

Class III-non-implantable

Conformity Assessment Route:

  • Annex IX, QMS – Chapters I, III
  • Technical documentation
  • Annex X – Type examination
  • Annex XI
  • Part A – Product Quality Assurance
  • Part B – Product Verification
  • Consultation if and when required – this will be determined on a case by case basis
  • Declaration of Conformity

2. Device Classification:

Class III Implantable

Conformity Assessment Route:

  • Annex IX, QMS – Chapters I & III
  • Technical documentation
  • Annex X – Type examination
  • Annex XI
  • Part A – Product Quality Assurance
  • Part B – Product Verification

We Can Help

Our MDR team is dealing with many clients and helping them to adjust to the new regulations. Talk to us today by calling 00353 52 61 76 706 or complete the form on our Contact Us page.

Read More

  More Blog Posts

Shortlisted for a Pharma Industry Award

Shortlisted for a Pharma Industry Award

We are thrilled that our Regulatory Team was shortlisted for a Pharma Industry Awards -Pharma Contract Services Company of the Year 2024. Our nomination is for their excellent work on implementing a client Regulatory Information Management (RIM) System.

read more
World Drug Safety Congress, Europe

World Drug Safety Congress, Europe

Our Head of PV is attending the World Drug Safety Congress during 9-10th October in Amsterdam this year. This is Europe's Most Exciting Drug Safety and PV event and with over 90+ exhibitors, 300+ speakers and 1,500 PV experts in attendance, it promises to be a truly...

read more

Strategy

Subscribe to Our Newsletter

Sign up for our newsletter today and stay informed on the latest regulatory updates and industry insights.